Implementation of Therapeutic and/or Abuse Drug Analysis in Blood Samples via Chemiluminescent Immunoassay Technique using the Randox Evidence Investigator™ Equipment

Authors

  • Dr. Mayra Judith Barrios L,

Keywords:

Drug, Blood, Immunoassay, chemiluminescence, biochip

Abstract

The Toxicology Laboratory of the Instituto Nacional de Ciencias Forenses de Guatemala -INACIF- is responsible for analysing the biological signs in the search for chemical substances, among which drugs are found. Because Laboratories constantly need to update their technology, it was necessary to implement a new chemiluminescence immunoassay technique using the Randox Evidence Investigator TM kit. This technique is useful as it provides accurate and reproducible results. To implement this technique, 10 blood samples with a positive result for drugs habitually detected in the Toxicology Laboratory and 10 blood samples with a negative result for drugs were analysed, said analysis gave as a result the fulfilment of what was expected for positive and negative drugs. The time invested in analysis was also evaluated, which was approximately 4 to 4.5 hours.

References

J. Bello, A. López (2001) Fundamentos de Ciencia Toxicológica. p. 18.

Congreso de la República de Guatemala (1992) Decreto 48-1992, Ley contra la Narcoactividad.

Randox Laboratories Ltd. (2016) Instructivo RANDOX Drogas matriz DOA I WB P.

Randox Laboratories Ltd. (2016) Manual del operador Evidence Investigator, V03.0.1.

Randox Laboratories Ltd. (2020) Validación RANDOX DOA ultra whole blood array (DOA ULTRA WB).

Randox Laboratories Ltd. (2022) Biochip Array Technology. https://www.randox.com/multiplex-biochip-testing/

D. Skoog, D. West (2015) Fundamentos de Química Analítica. p. 770.

E. Stashenko, J. R. Martínez (2012) GC-MS: herramienta fundamental para el análisis de drogas de uso ilícito. (4(1)), 21-33.

SWGTOX (2013) Standard Practices for Method Validation in Forensic Toxicology, Doc 003, revision 1.

Implementation of Therapeutic and/or Abuse Drug Analysis in Blood Samplesvia Chemiluminescent Immunoassay Technique using the Randox Evidence Investigator™ Equipment

Downloads

Published

2024-11-19

How to Cite

Implementation of Therapeutic and/or Abuse Drug Analysis in Blood Samples via Chemiluminescent Immunoassay Technique using the Randox Evidence Investigator™ Equipment. (2024). London Journal of Research In Science: Natural and Formal, 24(14), 31-37. https://www.journalspress.uk/index.php/LJRS/article/view/1471